Model | CS-3000HDR | CS-3000 | CS-2000Plus |
Measurement wavelength range | 380 to 780 nm |
Wavelength resolution | 0.9 nm/pixel |
Display wavelength interval | 1.0 nm |
Wavelength precision | ±0.3 nm (center-of-gravity wavelength mercury cadmium lamp: 435.8 nm, 546.1 nm, 643.8 nm) |
Spectrum wavelength width | 5 nm max. (half width value) |
Measurement angle | 1°, 0.2°, 0.1° |
Minimum measurement diameter | 1° | Ø5 mm (Ø1 mm with closeup lens) |
0.2° | Ø1 mm (Ø0.2 mm with closeup lens) |
0.1° | Ø0.5 mm (Ø0.1 mm with closeup lens) |
Minimum objective distance | 350 mm (55 mm with closeup lens) |
Minimum radiance display | 0.00002 cd/m2 |
Minimum spectral radiance display | 1.0 × 10–9 W/(sr·m2·nm) |
Luminance range with guaranteed accuracy (Light source A) | 1° | 0.0001 to 100,000 cd/m2 | 0.0005 to 5,000 cd/m2 | 0.003 to 5,000 cd/m2 |
0.2° | 0.0025 to 2,500,000 cd/m2 | 0.0125 to 125,000 cd/m2 | 0.075 to 125,000 cd/m2 |
0.1° | 0.01 to 10,000,000 cd/m2 | 0.05 to 500,000 cd/m2 | 0.3 to 500,000 cd/m2 |
Luminance
(Light source A) | ACCURACY*1 | ±5% (0.0001 to 0.0004 cd/m2) | ±2% | ±2% |
±2% (0.0004 to 10,000,000 cd/m2) |
REPEATABILITY (2Σ)*2 | 1° | ≥ 0.0001 cd/m2 | 5% | - | - |
≥ 0.0004 cd/m2 | 1.50% | 1.5% (≥ 0.0005 cd/m2) | - |
≥ 0.001 cd/m2 | 0.70% | 0.70% | - |
≥ 0.003 cd/m2 | 0.25% | 0.25% | 0.40% |
≥ 0.05 cd/m2 | 0.15% | 0.15% | 0.3% (≥ 0.05 cd/m2) |
0.2° | ≥ 0.0025 cd/m2 | 5% | - | - |
≥ 0.01 cd/m2 | 1.50% | 1.5% (≥ 0.0125 cd/m2) | - |
≥ 0.025 cd/m2 | 0.70% | 0.70% | - |
≥ 0.075 cd/m2 | 0.25% | 0.25% | 0.40% |
≥ 1.25 cd/m2 | 0.15% | 0.15% | 0.3% (≥ 1.25 cd/m2) |
0.1° | ≥ 0.01 cd/m2 | 5% | - | - |
≥ 0.04 cd/m2 | 1.50% | 1.5% (≥ 0.05 cd/m2) | - |
≥ 0.1 cd/m2 | 0.70% | 0.70% | - |
≥ 0.3 cd/m2 | 0.25% | 0.25% | 0.40% |
≥ 5 cd/m2 | 0.15% | 0.15% | 0.3% (≥ 5 cd/m2) |
|
| 0.15% (≥ 10 cd/m2) |
Chromaticity (Light source A) | ACCURACY*1 | 1° | ≥ 0.001 cd/m2 | x: ±0.002 | y: ±0.002 | x: ±0.002 | y: ±0.002 | x: ±0.003 | y: ±0.003 |
u': ±0.0022 | v': ±0.0011 | u': ±0.0022 | v': ±0.0011 | u’: ±0.0033 | v’: ±0.0016 |
|
| x: ±0.002 | y: ±0.002 |
u’: ±0.0022 | v’: ±0.0011 |
≥ 0.05 cd/m2 | x: ±0.0015 | y: ±0.001 | x: ±0.0015 | y: ±0.001 | x: ±0.0015 | y: ±0.001 |
u': ±0.0014 | v': ±0.0006 | u': ±0.0014 | v': ±0.0006 | u': ±0.0014 | v': ±0.0006 |
0.2° | ≥ 0.025 cd/m2 | x: ±0.002 | y: ±0.002 | x: ±0.002 | y: ±0.002 | x: ±0.003 | y: ±0.003 |
u': ±0.0022 | v': ±0.0011 | u': ±0.0022 | v': ±0.0011 | u’: ±0.0033 | v’: ±0.0016 |
|
| x: ±0.002 | y: ±0.002 |
u’: ±0.0022 | v’: ±0.0011 |
≥ 1.25 cd/m2 | x: ±0.0015 | y: ±0.001 | x: ±0.0015 | y: ±0.001 | x: ±0.0015 | y: ±0.001 |
u': ±0.0014 | v': ±0.0006 | u': ±0.0014 | v': ±0.0006 | u': ±0.0014 | v': ±0.0006 |
0.1° | ≥ 0.1 cd/m2 | x: ±0.002 | y: ±0.002 | x: ±0.002 | y: ±0.002 | x: ±0.003 | y: ±0.003 |
u': ±0.0022 | v': ±0.0011 | u': ±0.0022 | v': ±0.0011 | u’: ±0.0033 | v’: ±0.0016 |
|
| x: ±0.002 | y: ±0.002 |
u’: ±0.0022 | v’: ±0.0011 |
≥ 5 cd/m2 | x: ±0.0015 | y: ±0.001 | x: ±0.0015 | y: ±0.001 | x: ±0.0015 | y: ±0.001 |
u': ±0.0014 | v': ±0.0006 | u': ±0.0014 | v': ±0.0006 | u': ±0.0014 | v': ±0.0006 |
REPEATABILITY (2Σ)*2 | 1° | ≥ 0.001 cd/m2 | x: 0.0030 | y: 0.0035 | x: 0.0030 | y: 0.0035 | - |
u': 0.0024 | v': 0.0014 | u': 0.0024 | v': 0.0014 |
≥ 0.003 cd/m2 | x: 0.0010 | y: 0.0015 | x: 0.0010 | y: 0.0015 | x: 0.002 | y: 0.002 |
u': 0.0009 | v': 0.0006 | u': 0.0009 | v': 0.0006 | u': 0.0016 | v': 0.0008 |
|
| x: 0.001 | y: 0.001 |
u’: 0.0008 | v’: 0.0004 |
≥ 0.1 cd/m2 | x: 0.0006 | y: 0.0006 | x: 0.0006 | y: 0.0006 | x: 0.0006 | y: 0.0006 |
u': 0.0005 | v': 0.0002 | u': 0.0005 | v': 0.0002 | u': 0.0005 | v': 0.0002 |
≥ 0.2 cd/m2 | x: 0.0004 | y: 0.0004 | x: 0.0004 | y: 0.0004 | x: 0.0004 | y: 0.0004 |
u': 0.0003 | v': 0.0002 | u': 0.0003 | v': 0.0002 | u': 0.0003 | v': 0.0002 |
0.2° | ≥ 0.025 cd/m2 | x: 0.0030 | y: 0.0035 | x: 0.0030 | y: 0.0035 | - |
u': 0.0024 | v': 0.0014 | u': 0.0024 | v': 0.0014 |
≥ 0.075 cd/m2 | x: 0.0010 | y: 0.0015 | x: 0.0010 | y: 0.0015 | x: 0.002 | y: 0.002 |
u': 0.0009 | v': 0.0006 | u': 0.0009 | v': 0.0006 | u': 0.0016 | v': 0.0008 |
|
| x: 0.001 | y: 0.001 |
u’: 0.0008 | v’: 0.0004 |
≥ 2.5 cd/m2 | x: 0.0006 | y: 0.0006 | x: 0.0006 | y: 0.0006 | x: 0.0006 | y: 0.0006 |
u': 0.0005 | v': 0.0002 | u': 0.0005 | v': 0.0002 | u': 0.0005 | v': 0.0002 |
≥ 5 cd/m2 | x: 0.0004 | y: 0.0004 | x: 0.0004 | y: 0.0004 | x: 0.0004 | y: 0.0004 |
u': 0.0003 | v': 0.0002 | u': 0.0003 | v': 0.0002 | u': 0.0003 | v': 0.0002 |
0.1° | ≥ 0.1 cd/m2 | x: 0.0030 | y: 0.0035 | x: 0.0030 | y: 0.0035 | - |
u': 0.0024 | v': 0.0014 | u': 0.0024 | v': 0.0014 |
≥ 0.3 cd/m2 | x: 0.0010 | y: 0.0015 | x: 0.0010 | y: 0.0015 | x: 0.002 | y: 0.002 |
u': 0.0009 | v': 0.0006 | u': 0.0009 | v': 0.0006 | u': 0.0016 | v': 0.0008 |
|
| x: 0.001 | y: 0.001 |
u’: 0.0008 | v’: 0.0004 |
≥ 10 cd/m2 | x: 0.0006 | y: 0.0006 | x: 0.0006 | y: 0.0006 | x: 0.0006 | y: 0.0006 |
u': 0.0005 | v': 0.0002 | u': 0.0005 | v': 0.0002 | u': 0.0005 | v': 0.0002 |
≥ 20 cd/m2 | x: 0.0004 | y: 0.0004 | x: 0.0004 | y: 0.0004 | x: 0.0004 | y: 0.0004 |
u': 0.0003 | v': 0.0002 | u': 0.0003 | v': 0.0002 | u': 0.0003 | v': 0.0002 |
Polarization error | 2% max. (400 to 780 nm) at 1° setting; 3% max. (400 to 780 nm) at 0.2° or 0.1° setting |
Integration time | 0.005 to 92 seconds (Normal mode) |
| 0.005 to 120 seconds (Normal mode) |
0.005 to 16 seconds (Fast mode) |
| 0.005 to 16 seconds (Fast mode) |
Measurement time | STANDALONE | Minimum of 1 second or less (Manual mode) to approx. 190 seconds (Normal mode) or maximum of approx. 242 seconds (Manual mode) |
| Minimum of 1 second or less (Manual mode) to maximum of approx. 242 seconds (Normal or Manual mode) |
COMMUNICATION*3 | CS-3000HDR, CS-3000: Approx. 0.07 second |
CS-2000Plus: Approx. 0.08 second |
(when Manual mode, 33.333 ms and Intelligent Dark function are set) |
Colour modes | Lvxy, Lvu’v’, LvTΔuv, XYZ, spectrograph, main wavelength, excitation purity |
Colour matching functions | 2° field of view, 10° field of view, CIE 170-2:2015 (2°, 10°), user-specified colour matching functions (supported by measurement software) |
Function for detecting/setting optical frequency of measurement objective | Yes | No |
Measurement angle switching | Powered | Manual |
Interfaces | USB 2.0; RS-232C |
Operating temperature and humidity range | 5 to 30°C, 80% RH max. (no condensation) | 5 to 35°C, 80% RH max. (no condensation) |
Storage temperature and humidity range | 0 to 35°C, 80% RH max. (no condensation) |
Power supply | Dedicated AC adapter (100 to 240 V, 50/60 Hz) |
Power consumption | Approx. 20 W |
Size | Spectroradiometer: 158 × 262 × 392 (W × H × D) mm |
Weight | Approx. 7.0 kg |